Chapter 3, Description of Screens
Call Control Screens
The messages which can be viewed are:
Reverse Control Channel Messages for Paging, Origination, Orders, and Order
Confirmation.
RECCW A - Word A - Abbreviated Address Word
RECCW B - Word B - Extended Address Word
RECCW C - Word C - Serial Number Word
RECCW D - Word D - First Word of the Called-Address
RECCW E - Word E - Second Word of the Called-Address
Reverse Voice Channel Messages for Order Confirmation.
RVCOrdCon - Reverse Voice Channel Order Confirmation Message
Reverse Control Channel Authentication Messages.
BSChalOrd - Word C - Base Station Challenge Word
AuthWORD - Word C - Authentication Word
UniqChCon - Word C - Unique Challenge Order Confirmation Word
Reverse Voice Channel Authentication Messages.
RVCOrd - Word one - Reverse Voice Channel Generic Order
RVCBSChal - Word 2 - Reverse Voice Channel Base Station Challenge Order
NAMPS Reverse Voice Channel Messages.
NRVC Ord - Narrow Order or Order Confirmation Message
MRI Ord - MRI Order message
The field descriptions for each of the decoded messages are given in the Display
Word
field description, on page 232.
When the CALL DATA screen is displayed and the Call Processing Subsystem is
in the connected state, the Test Set constantly monitors the mobile station’s
transmitted carrier power. If the power falls below 0.0005Watts, the error message
RF Power Loss indicates loss of Voice Channel will be displayed
and the Test Set will terminate the call and return to the active state.
NOTE: In order to ensure that the Test Set makes the correct decisions regarding the presence of the
mobile stations’s RF carrier, the Test Set’s RF power meter should be zeroed before using the
Call Processing Subsystem. Failure to zero the power meter can result in erroneous RF power
measurements.