251
Chapter 4, Description of Fields
Fields
Execute (Closed Loop Power Control Change)
This field allows execution of n-up, n-down, or ramp power-control steps.
GPIB Example
“DISP CTXR;CDMA:PCONtrol:EXECute”
causes the Test Set to output the selected number of steps in the direction selected
in the
Change field over the power control sub-channel.
Operating Considerations
When Execute is selected, the Test Set will begin sending power-control bits to
the mobile station. Each power-control bit attempts to adjust the mobile station’s
power level by 1 dB.
See Also
Change
field description, on page 193
Steps
field description, on page 414
Screens Where Field is Present
CDMA TRANSMITTER POWER RANGE TEST
Exec (Execution Cond)
Selecting this field displays the TESTS (Execution Conditions) screen.
Screens Where Field is Present
TESTS (Main Menu)
Expected
This column in the authentication results table lists Test Set’s computed
authentication data for the parameters that are returned from the mobile in the
Access Channel Registration, Page Response Message, or Origination Message.
The data in the
Expected column is compared to the data in the Received
column to verify if authentication passes or fails.
See Also
Authen
field description, on page 170
Screens Where Field is Present
CDMA AUTHENTICATION