Chapter 4, Description of Fields
Fields
Operating Considerations
The Test Set measures the time response of the mobile station’s output power in
response to a step change in the Test Set’s output power. Test standards require
that sector A power (
"Sctr A Pwr" field on page 380) initially be set to -60 dBm.
The first step is +20 dB (to -40 dBm). The next step is -20 dB (back to -60 dBm).
The third step is another -20 dB (to -80 dBm), and the final step is +20 dB (back to
-60 dBm).
See Also
Open Loop Time Response (Execute)
field description, on page 320
Screens Where Field is Present
OPEN LOOP TIME RESPONSE
Step#
The Step# field lists the order of selected tests for the current procedure.
Operating Considerations
Use the knob, up-arrow, or down-arrow key to select the step where you want to
delete, replace, or insert a test.
Screens Where Field is Present
TESTS (Order of Tests)
Stop Length
This field specifies the number of stop bits used for serial communications when
using the rear-panel serial port. Selecting this field displays a list of stop bit
choices.
This setting is maintained after the instrument is turned off.
Screens Where Field is Present
I/O CONFIGURE