Agilent Technologies E8285A Video Games User Manual


 
54
S:\agilent\e8285\REFGUIDE\MANUAL\keys.chp
Chapter 1, Description of Keys
Keys
Register
This key causes the mobile station to perform a CDMA zone-based registration.
See
"Register (CDMA)" on page 358.
Release
This key is used to delete the assignment of the User Keys, K1 through K5, and
K1’ through K3’. Press and release the Shift key, then the K5 key to select the
Release function.
GPIB Example
No equivalent GPIB command
See Also
“Using USER Keys” in the Operating Overview chapter of the E8285A
User’s Guide
IBASIC User’s Handbook to assign IBASIC commands to User Keys
Save
Press and release the Shift key, then the Recall key to access the save function.
This stores instrument setups.
Programming Save
The GPIB commands :REGister:SAVE are used to select this function
programmatically. The save recall mass storage device is selected using the
Save/Recall field on the I/O CONFIGURE screen.
Example
OUTPUT 714;"REG:SAVE ’SETUP1’"
This saves the instrument state to a file named SETUP1:
Removing Saved Instrument States. One or all of the saved instrument states can
be removed from the selected save/recall mass storage device. The
save/recall mass storage device is selected using the
Save/Recall field on the
I/O CONFIGURE screen. The GPIB commands :REGister:CLEar are used to
perform this function programmatically.