Agilent Technologies E8285A Video Games User Manual


 
270
S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp
Chapter 4, Description of Fields
Fields
Grid (CDMA Swept Power Measurement)
This field allows the user to superimpose a grid over the display on the screen.
GPIB Example
“CDMA:SWEPtpower:GRID ‘On’”
sets the grid to on or off.
Operating Considerations
Select On to enable the grid; select Off to disable the grid. The default is On.
The dashed grid lines represent the time scale on the X axis and the relative level
scale on the Y axis.
Screens Where Field is Present
CDMA SWEPT POWER MEASUREMENT (When the Controls field is set to
Display)
Handoff
This field is used to initiate a handoff.
The voice channel number to hand the mobile station off to, the initial power level
to use on the new voice channel and the SAT tone frequency to transpond on the
new voice channel are specified using the
Chan:, Ch Loc:, Pwr Lvl:, DSAT:,
and
SAT: fields under the Voice Channel Assignment section of the CALL
CONTROL
screen.
The :HANDoff command is used to control this field.
There is no query form of the :HANDoff command.
Syntax
:HANDoff
Example
OUTPUT 714;"CALLP:HAND"
Screens Where Field is Present
CALL CONTROL