Agilent Technologies E8285A Video Games User Manual


 
47
Chapter 1, Description of Keys
Keys
To Set the Meter High End and Low End Points.
Use the :METer:HEND and :MEter:LEND commands to set the analog bar-graph
meter’s high endpoint and low endpoint.
Syntax
:METer:HEND <real number>
:METer:LEND <real number>
Example
OUTPUT 714;"MEAS:RFR:POW:MET:HEND 20"
OUTPUT 714;"MEAS:RFR:POW:MET:LEND 10"
This sets the analog bar-graph meter’s high endpoint to 20 watts and the low
endpoint to 10 watts for the TX power measurement.
NOTE: When setting the value of the Meter data function through GPIB, a
non-Attribute Unit unit-of-measure must be specified in the command string, otherwise the
current Attribute Unit is assumed by the Test Set.
Refer to “To Specify Units-of-Measure for GPIB Data Transfer” in
the Operating Overview
chapter of the E8285A User’s Guide.
To Query the Meter High End and Low End Points.
Use the :METer:HEND? and :MEter:LEND? commands to query the analog bar-
graph meter high endpoint and low endpoint.
Syntax
:METer:HEND?
:METer:LEND?
OUTPUT 714;"MEAS:RFR:POW:MET:HEND?"
ENTER 714;Meter_hi_end
OUTPUT 714;"MEAS:RFR:POW:MET:LEND?"
ENTER 714;Meter_lo_end
This queries the high end point and low end point of the analog bar-graph
meter for the TX power measurement.
NOTE: When querying the value of the Meter data function through GPIB, the Test Set always returns
numeric values in attribute units, regardless of the current display units or GPIB units settings.
Numeric values are expressed in scientific notation.
Refer to “To Specify Units-of-Measure for GPIB Data Transfer” in the Operating Overview
chapter of the E8285A User’s Guide.