Chapter 4, Description of Fields
Fields
Max Req Seq, Max Rsp Seq
This field allows simultaneous entry of the maximum number of access probe
sequences for an access channel request and the maximum number of access
probe sequences for an access channel response (MAX_REQ_SEQ or
MAX_RSP_SEQ). These are separate parameters in the CDMA specification but
the Test Set sets both parameters to a single value.
GPIB Example
“DISP CCON;CDMA:CELL:CONFigure:MAXReqrspseq 3“
sets both parameters to 3.
Operating Considerations
This field indicates the maximum number of access probe sequences sent by the
mobile when the mobile is requesting service (Max Req Seq) or when the mobile
has been paged (Max Rsp Seq). If the Call Limit field on the CDMA Call Site
Configuration screen is set to “Page”, the mobile station will transmit the number
of access probe sequences entered in this field.
Screens Where Field is Present
CDMA CELL SITE CONFIGURATION
Max Slot Cycle Index
Max Slot Cycle Index allows you to set the slot cycle used by the Test Set to page
a mobile station. It is set by integer values, which correspond to time using the
following formula: seconds, where i has integer values from 0 to
7 (0 is default).
The Test Set will use the lesser of the value entered in this field and the slot cycle
index of the mobile station when the MS ID field is set to Auto mode and the
mobile station has registered.
NOTE When i = 7, the value is quite large at 163.84 seconds. You may not want to wait
this long during normal testing.
GPIB Example
“CDMA:CELL:CONFigure:MSCindex 1”
sets the cycle to 2.56 seconds.
Screens Where Field is Present
CDMA CELL CONFIGURE
T 1.28 2