Chapter 4, Description of Fields
Fields
Sensitivity
RF Analyzer sensitivity adds about 6 dB of sensitivity for the ANTENNA IN port
when
High is selected.
Operating Considerations
Selecting High sensitivity may cause Spectrum Analyzer measurements to be
uncalibrated when the ANTENNA IN port is used (a message appears when this
occurs).
High-level AM measurements may be distorted when high sensitivity is used with
the ANTENNA IN port.
Screens Where Field is Present
RF ANALYZER
Sensitivity
This area performs two functions:
The upper field selects Normal or High sensitivity for the RF input. The High setting
adds about 6 dB of sensitivity to the ANTENNA IN port for looking at very low level
signals. However, this setting can cause measurements to be uncalibrated (indicated by
a message on the screen). High sensitivity can also cause high-level AM signals to be
distorted.
The lower field selects the vertical resolution of the display. You can choose from
1 dB, 2 dB, or 10 dB per graticule.
Screens Where Field is Present
SPECTRUM ANALZYER
CDMA REVERSE CHANNEL SPECTRUM
Serial Baud
This field selects the baud rate for serial communications when using the rear-
panel serial port. Selecting this field displays a list of baud rate choices.
This setting is maintained after the instrument is turned off.
Screens Where Field is Present
I/O CONFIGURE