Chapter 4, Description of Fields
Fields
Num Pages
This field allows entry of the number of page attempts the Test Set will perform
when a call is made from the Test Set. The power-up or preset default value is one
page attempt.
GPIB Example
“CDMA:CELL:CONF:NPAG 2“
sets the Num Pages field to two.
Operating Considerations
Multiple pages (two or three) should only be selected when testing non-slotted-
mode mobile stations. If multiple pages are selected when testing a slotted-mode
mobile station, connect time could take minutes.
Screens Where Field is Present
CDMA CELL SITE CONFIGURATION
Num Step
This field allows entry of the number of power steps in an access probe sequence
(NUM_STEP).
GPIB Example
“DISP CCON;CDMA:CELL:CONFigure:NUMStep 0“
sets NUM_STEP to 0.
Operating Considerations
This field allows the Test Set to set the maximum number of access probes the
mobile-station-under-test will transmit in each access probe sequence. The
number of probes transmitted will be 1 + NUM_STEP.
Screens Where Field is Present
CDMA CELL SITE CONFIGURATION