Chapter 4, Description of Fields
Fields
Marker Pos (CDMA Reverse Channel Spectrum)
This field is displayed when the Controls field is set to Marker. This field sets
the marker’s position, referenced to the left side of the display.
GPIB Example
“DISP CSP;CSP:MARKer:POSition 5.0“
positions the marker in the center of the display.
Operating Considerations
The position is settable in units of divisions.
Screens Where Field is Present
CDMA REVERSE CHANNEL SPECTRUM
Marker Time (CDMA Gated Power)
This field is displayed when the Controls field on the CDMA GATED POWER
screen is set to
Display. This field sets the marker’s position in microseconds,
referenced to the time alignment of the 1.25 ms power control groups in the gated
power measurement ensemble average.
GPIB Example
“CDMA:GPOWer:MARKer:TIME 0“
positions the marker at 0 microseconds.
“MEAS:GPOWer:MARKer:LEVel?“
queries the gated power ensemble average level at the current marker position.
Operating Considerations
The position is settable in units of microseconds.
Screens Where Field is Present
CDMA GATED POWER