323
Chapter 4, Description of Fields
Fields
Output Atten Hold
This field, when set to On, prevents the output attenuators in the Test Set from
switching in and out, avoiding temporary loss of the output signal as the RF
Power level is changed.
On
Off
GPIB Example
“RFG:ATT ‘On’”
holds the output attenuators.
Operating Considerations
Always adjust the RF Power level (the sum of Sctr A Pwr, Sctr B Pwr, and
AWG N) to the highest power level in the range of values you will be testing
before turning Output Atten Hold “On”.
If the RF Power is raised (or lowered) beyond the limits imposed by attenuator
settings being held, the Test Set will beep and the error message “Change RF Gen
Amplitude, Output Port or Atten Hold (if on)” will be displayed.
This field is coupled to the Atten Hold field on the analog RX TEST, RF GEN,
and DUPLEX TEST screens.
Screens Where Field is Present
CDMA TRANSMITTER POWER RANGE TEST
Output Heading:
This field allows you to enter a heading for the test results printout (or CRT
display).
Screens Where Field is Present
TESTS (Printer Setup, Execution Conditions)