Chapter 4, Description of Fields
Fields
Mask Type
This field is used to select the power versus time mask used during gated power
measurements to determine pass/fail results.
Original selects the mask specified by test standards IS-98A and TIA/EIA-98B.
TIA/EIA-98-C selects the mask specified by test standard TIA/EIA-98C.
Screens Where Field is Present
CDMA GATED POWER (when Display is selected from the Controls field.)
Max Frames
This field sets the maximum frame count for each FER test.
GPIB Example
“DISP CRXT;MEASure:CDManalyzer:FERasure:MAXimum:FRAMes 1000”
sets the maximum number of frames for FER measurements to 1000.
Operating Considerations
FER measurements will be made on the number of frames entered in this field
unless the test terminates because a confidence limit is reached.
Range of values: 25 through 10000000
See Also
Confidence
field description, on page 207
Screens Where Field is Present
CDMA CELLULAR MOBILE RECEIVER TEST