Chapter 4, Description of Fields
Fields
Nom Power
This field allows entry of the nominal transmit power offset (NOM_PWR).
GPIB Example
“DISP CCON;CDMA:CELL:CONFigure:NOMinal:POWer 0“
sets NOM_PWR to 0.
Operating Considerations
Nom Power is an adjustment to the mobile station’s open loop power level. The
mobile station obtains this value prior to transmitting, as part of the Access
Parameters Message.
Screens Where Field is Present
CDMA CELL SITE CONFIGURATION
Nom Pwr Ext
This field allows entry of the extended nominal transmit power
(NOM_PWR_EXT). This field is displayed only when the field
"Protocol" on
page 346
is set to J-STD-008 or TIA/EIA-95B.
GPIB Example
“CDMA:CELL:CONFigure:NOMinal:POWer:EXT 0“
sets NOM_PWR_EXT to 0.
Operating Considerations
Entering a 0 in this field sets the range of the open loop power control correction
factor from
8 dB to 7 dB inclusive. Entering a 1 extends the correction factor
from
24 to 9 dB inclusive. The mobile station obtains this value prior to
transmitting, as part of the Access Parameters Message.
Screens Where Field is Present
CDMA CELL SITE CONFIGURATION