Agilent Technologies E8285A Video Games User Manual


 
212
S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp
Chapter 4, Description of Fields
Fields
Controls (CDMA Swept Power Measurement)
This field allows the user to select which of the CDMA swept power measurement
menus will be displayed.
GPIB Example
“CDMA:SWEPtpower:CONTrol ’Main’“
selects Main from the CDMA SWEPT POWER MEASUREMENT screen Controls
field.
Operating Considerations
Select among the following:
Main displays Swept Pwr and Ref Level.
Trigger displays Amplitude and Slope.
Display displays Marker Time, Trace, and Grid.
Auxiliary displays RF Channel and Input Atten.
The default selection is Main.
Screens Where Field is Present
CDMA SWEPT POWER MEASUREMENT
Controls (Open Loop Time Response)
This field allows the user to select which of the open loop time response
measurement menus is displayed.
Main
Display
CDMA Gen
Traffic
GPIB Example
“CDMA:OLTR:CONTrol ‘Main’“
selects “Main” from the OPEN LOOP TIME RESPONSE screen Controls field.
Screens Where Field is Present
OPEN LOOP TIME RESPONSE