Chapter 4, Description of Fields
Fields
RANDU
RANDU is the 24-bit random number issued by the base station in the Unique
Challenge Order. RANDU is entered as a 6-digit hexadecimal value. RANDU is
user definable.
GPIB Example
“CDMA:CELL:CONF:AUTH:RANDU?”
queries the RANDU field
See Also
"Uniq Chall" field on page 441
Screens Where Field is Present
CDMA AUTHENTICATION
Range Hold
These fields enable/disable several autoranging and autotuning routines.
Auto All enables these routines, providing automatic adjustment when making
AF or RF measurements.
Hold All disables these routines, requiring you to manually set the affected
settings.
The following fields are affected by the
Range Hold field:
RX/TX Cntl in the CONFIGURE screen.
Input Atten in the RF ANALYZER and SPECTRUM ANALYZER screens.
Gain Cntl in the AF ANALYZER screen. This field controls three AF gain setting
fields:
Input Gain
De-Emp Gain
Notch Gain
Operating Considerations
The Hold All setting is primarily used when the instrument is operated by
remote control, such as in an automated test system.
Unless you have very specific reasons for disabling the automatic functions, you
should set this field to
Auto All when operating the instrument manually.