Agilent Technologies E8285A Video Games User Manual


 
433
Chapter 4, Description of Fields
Fields
Traffic Data Mode
This field selects the desired service option. “Rate set 2” refers to vocoders that
operate at 14.4 kips full rate.
Svc Opt 1 - normal traffic
Svc Opt 2 - data loopback
Svc Opt 3 - service redirection
Svc Opt 6 - short message service (for rate set 1)
Svc Opt 9
- data loopback (for rate set 2 vocoders, displayed when the
Protocol field is set to TSB-74 or J-STD-008)
Svc Opt 14 - short message service (for rate set 2)
Svc Opt 32768
- normal traffic (for rate set 2 vocoders, displayed when the
Protocol field is set to TSB-74 or J-STD-008)
GPIB Example
“DISP CCNT;CDMA:CALL:TRAFfic:DATA:MODE ‘SVC OPT 1’“
selects service option 1.
Operating Considerations
Data loopback is used during most mobile station tests.
The desired service option can be selected at any time. If a service option change
is attempted while a call is connected, and the mobile station does not respond to
the message, a beep will sound and the service option will not change.
This field is duplicated on the CDMA CELLULAR MOBILE TRANSMITTER
TEST, CDMA CALL CONTROL, CDMA CELLULAR MOBILE RECEIVER
TEST, CDMA AUTHENTICATION, and CDMA SHORT MESSAGE SERVICE
screens.
Screens Where Field is Present
CDMA CALL CONTROL
CDMA TRANSMITTER POWER RANGE TEST
CDMA CELLULAR MOBILE RECEIVER TEST
CDMA CELLULAR MOBILE TRANSMITTER TEST
CDMA SHORT MESSAGE SERVICE
CDMA AUTHENTICATION