Chapter 4, Description of Fields
Fields
Release
This field is used to terminate an active voice channel connection with the mobile
station. When the
Release field is selected, a mobile station control message
with a release order is sent to the mobile station on the forward voice channel. A
mobile station must be actively connected on a voice channel to the Test Set (that
is, the
Connect annunciator must be lit) before you can attempt to send a release
order to the mobile station.
The :RELease command is used to control this field.
There is no query form of the :RELease command.
Syntax
:RELease
Example
OUTPUT 714;"CALLP:REL"
Screens Where Field is Present
CALL CONTROL
Res BW
The resolution bandwidth field selects the IF filter used when measuring the
power of the carrier and the adjacent channels.
Operating Considerations
Using a narrower bandwidth filter (300 Hz) slows the measurement, but rejects
carrier leakage and out of channel spurs. Using a wider bandwidth filter
(1 kHz) speeds measurements, but may allow unwanted spurs and carrier leakage
to be integrated into the measurement when measuring at the edges of the selected
channel bandwidth.
Screens Where Field is Present
ADJACENT CHANNEL POWER
Reset
See the Cont/Single field description, on page 211
Screens Where Field is Present
OSCILLOSCOPE