Operating Considerations for the Channel Power Measurement
The cross-calibration process is performed by selecting the Power Meas,
Calibrate
field. Refer to “Calibrating the Test Set” in the Calibrating the Test
Set chapter of the E8285A Application Guide. Channel power measurements must
be calibrated whenever any of the following conditions occur:
Uncal annunciator is flashing
5 °C change in operating temperature
Test Set power cycle
Daily
The Average Power measurement, which is used during cross-calibration, is
automatically zeroed during the calibration procedure.
No special measurement triggers are required for Channel Power measurements.
See Also
Avg Power
field description, on page 178
Screens Where Field is Present
CDMA CALL CONTROL
CDMA TRANSMITTER POWER RANGE TEST
CDMA CELLULAR MOBILE TRANSMITER TEST
Chan Power Meas Intrvl
This field allows you to set the interval for the Channel Power measurement. This
setting is only valid with Service Option 2 or Service Option 9, and during Full
Rate.
GPIB Examples
“CDManalyzer:CHANpower:INTerval 1”
sets the measurement interval to 1 ms.
Screens Where Field is Present
CDMA TX