Agilent Technologies E8285A Video Games User Manual


 
324
S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp
Chapter 4, Description of Fields
Fields
Output Port
This field selects the RF output port on the Test Set.
RF Out selects the RF IN/OUT port on the Test Set’s front panel and displays the RF
Out field for entering CDMA forward channel path loss.
Dupl selects the DUPLEX OUT port on the Test Set’s front panel and displays the
Duplex Out field for entering CDMA forward channel path loss.
GPIB Example
“CONF:OUTP ‘Rf Out’”
selects the RF Output port.
Operating Considerations
Maximum signal levels at each port are printed on the front panel. The Duplex
Out port provides higher power.
CAUTION: Applying reverse RF power to the DUPLEX OUT connector can damage the instrument. (A
message is displayed when an overpower conditions occurs.) Whenever possible, use the RF IN/
OUT connector when testing transceivers to prevent damage from accidental transmitter
keying.
If a reverse-power condition triggers the internal protection circuit, remove the reverse-power
signal and press the Meas reset key or turn the Test Set off and on to reset it.
Screens Where Field is Present
CONFIGURE
CDMA REVERSE CHANNEL SPECTRUM
SPECTRUM ANALYZER
Output Results For
This field selects which test results are output. The underlined entry is the active
choice.
All outputs all test results on to the location selected in the Output Results To:
field.
Failures outputs only the results of tests which fail.
Screens Where Field is Present
TESTS (Execution Conditions, Printer Setup)