Agilent Technologies E8285A Video Games User Manual


 
44
S:\agilent\e8285\REFGUIDE\MANUAL\keys.chp
Chapter 1, Description of Keys
Keys
To Reset Measurement Limit Detection. Use the :HLIMit:RESet and
:LLIMit:RESet commands to reset measurement limit detection. Once a high or
low measurement limit has been exceeded (:HLIMit:EXCeeded? returns a 1 or
:LLIMit:EXCeeded? returns a 1), measurement limit detection is disabled until
reset by the :RESet command.
Syntax
:HLIMit:RESet
:LLIMit:RESet
Example
OUTPUT 714;"MEAS:AFR:FM:HLIM:RES"
OUTPUT 714;"MEAS:AFR:FM:LLIM:RES"
This resets high and low measurement limit detection for the AF Analyzer FM de-
viation measurement.
See Also
“Setting Measurement Limits” in the Operating Overview chapter of the E8285A
User’s Guide.
Meas reset
Meas reset clears the measurement history for all of the instrument’s
measurement algorithms (such as the averaging and peak hold functions) to restart
all measurements that are in progress.
Programming Meas reset
The GPIB commands :MEASure:RESet are used to select this function
programmatically.
Example
OUTPUT 714;":MEAS:RES"
This resets all of the active measurements in the Test Set.