Agilent Technologies E8285A Video Games User Manual


 
290
S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp
Chapter 4, Description of Fields
Fields
Marker Time (Open Loop Time Response)
This field is displayed when the Controls field on the OPEN LOOP TIME
RESPONSE screen is set to
Display. This field sets the marker’s position in
milliseconds, referenced to the time alignment of the step change in the Test Set’s
output power.
GPIB Example
“CDMA:OLTR:MARKer:TIME 0“
positions the marker at 0 milliseconds.
“MEASure:OLTR:MARKer:LEVel?“
queries the power level of the last time response to open loop power control test at
the current marker position.
Operating Considerations
The position is settable in units of milliseconds.
Screens Where Field is Present
OPEN LOOP TIME RESPONSE
Marker Time (CDMA Swept Power Measurement)
This field sets the position of the measurement marker on the display.
GPIB Example
“CDMA:SWEPtpower:MARKer:TIME ‘30.0’”
sets the marker position at 30 milliseconds.
Operating Considerations
Adjust the position of the marker on the display as appropriate. This field allows
adjustment of the marker position over the full range of the trace, which is from
0.0 millisecond to 100 milliseconds. The default is 0.0 millisecond.
Screens Where Field is Present
CDMA SWEPT POWER MEASUREMENT (When the Controls field is set to
Display)