Chapter 4, Description of Fields
Fields
Open Loop Time Response (Execute)
This field is displayed on the OPEN LOOP TIME RESPONSE screen when
“Main” is selected in the
"Controls (Open Loop Time Response)" field on page 212.
Selecting this field executes a time response to open loop power control test.
GPIB Example
“MEAS:CDManalyzer:OLTR:EXECute”
executes a time response to open loop power control test.
Operating Considerations
When Execute is selected, the Test Set will step its output power up or down
depending on the value entered in the
"Step Size" field on page 414. The mobile
station’s output power as a function of time will be graphically displayed on the
OPEN LOOP TIME RESPONSE screen.
Screens Where Field is Present
OPEN LOOP TIME RESPONSE
Options
This field is used to enter the instrument’s option number or numbers if any.
Operating Considerations
This field may be left blank, or otherwise may include other calling name options,
for example:
Printer options – LN=#,START,END
Where # is the number of lines on each page.
Where START causes a form feed at the start of each printout.
Where END causes a form feed at the end of each printout.
Data collection options – NN
Where NN is the number of records (file size) for the mass-storage location where
data will be collected on disk or memory card. The default record size is 80.
Screens Where Field is Present
TESTS (External Devices)