Agilent Technologies E8285A Video Games User Manual


 
204
S:\agilent\e8285\REFGUIDE\MANUAL\fields.chp
Chapter 4, Description of Fields
Fields
Screens Where Field is Present
CDMA TRANSMITTER POWER RANGE TEST
Clr Scr
This field is used to clear the IBASIC controller screen.
Screens Where Field is Present
TESTS (IBASIC Controller)
CMAX
The CMAX field sets the number of access channels in the system. This will
determine how many channels must be scanned by the mobile station when trying
to access the Test Set. The value of this field will affect the time required for the
mobile station to connect with the Test Set.
The :CMAXimum command is used to control this field.
The
CMAX field is an immediate action field. That is, whenever the :CMAXimum
command is sent, the change is reflected immediately in the appropriate signaling
message(s) being sent on the forward control channel. No change occurs to the
current state (i.e. Active, Register, Page, Access, Connect) of the Call Processing
Subsystem.
The query form of the command (that is, :CMAXimum?) can be used to
determine the current control channel setting.
Syntax
:CMAXimum <integer number>
:CMAXimum?
Example
OUTPUT 714;"CALLP:CMAX 21"
OUTPUT 714;"CALLP:CMAX?"
ENTER 714;Num_acc_chans
Screens Where Field is Present
CALL CONFIGURE