Chapter 4, Description of Fields
Fields
Input Att
Input Att sets the amount of input attenuation for the RF IN/OUT and ANTENNA
IN connectors. The Input Att field displays the present attenuation value, and it is
used to set the desired attenuation level.
The :CRFAttenuator command is used to control this field.
The query form (that is, :CRFAttenuator?) can be used to query the current field
value.
Syntax
:CRFAttenuator <‘><0 dB/20 dB/40 dB><’>
:CRFAttenuator?
Example
OUTPUT 714;"CALLP:AMPS:CRFA ‘40 dB’"
OUTPUT 714;"CALLP:AMPS:CRFA?"
ENTER 714;Attenuator_set$
Operating Considerations
This is a priority control field. Accessing the CALL CONFIGURE screen
overrides any changes made to this field in other screens. The other screens
affected include the RF ANALYZER screen and the SPECTRUM ANALYZER
screen.
Screens Where Field is Present
CALL CONTROL
CALL CONFIGURE