Chapter 4, Description of Fields
Fields
Power Cntl Step Size
This field is displayed only when the "Protocol" field on page 346 is set to TIA/
EIA-95B.
This field allows entry of the following closed loop power control step sizes:
1 dB
0.5 dB
0.25 dB
Operating Considerations,
The power control step size entered in this field must not be less than the
minimum step size supported by the mobile station under test.
The value entered in this field will be sent to the mobile station in the Power
Control Message PWR_CNTL_STEP field during call setup or while on a traffic
channel (call connected).
The mobile station’s minimum step size, MIN_PWR_CNTL_STEP, is retrieved
by the Test Set during registration and displayed in the
"MS Database" field on
page 306
under Pwr Step.
The message “Mobile’s minimum power step > selected Power Cntl Step Size”
will be displayed if a value is entered into the Power Cntl Step Size field that the
Test Set determines is unacceptable
See Also
Protocol
field description, on page 346
"MS Database" field on page 306
Screens Where Field is Present
CDMA TRANSMITTER POWER RANGE TEST