Agilent Technologies E8285A Video Games User Manual


 
34
S:\agilent\e8285\REFGUIDE\MANUAL\keys.chp
Chapter 1, Description of Keys
Keys
To Query the Number of Averages. Use the :AVERage:VALue? commands to
query the number of averages used by the averaging algorithm.
Syntax
:AVERage:VALue?
Example
OUTPUT 714;"MEAS:AFR:DISTN:AVER:VAL?"
ENTER 714;Num_of_avgs
This queries the number of averages for the Average data function for the AF An-
alyzer distortion measurement.
See Also
“To Use Measurement Averaging” in the Operating Overview chapter of the
E8285A User’s Guide.
Call/Page
When this key is pressed, the Test Set attempts a CDMA-mode page to a mobile
station. The Call Status annunciators on the CDMA CALL CONTROL screen
indicate call flow.
Refer to “Setting Up a Call” chapter in the E8285A Application Guide.
GPIB Example
“DISP CCNT;CDMA:CALL:MAKE”
Cancel
Cancel is used to cancel an entry in progress, or to stop a running IBASIC
program. For example, if you press Recall to recall an instrument setup, and then
decide not to recall a setting, pressing Cancel exits the recall procedure.
GPIB Example
No equivalent GPIB command
EEX
Press and release the Shift key, then the +/- key to access the exponent function.
This function is used for entering numbers using scientific notation.
See Also
“To Enter Values with Exponents” in the Operating Overview chapter of the
E8285A User’s Guide