Chapter 4, Description of Fields
Fields
Disarm
This field stops a measurement in progress. It is only displayed when the Meas
Cntl
field is set to Single. This field is not updated to show the Meas Cntl mode
when the Test Set is operated remotely.
GPIB Example
“DISP CRXT;TRIGger:ASTate ’DISARM’“
stops a measurement in progress.
Operating Considerations
The ‘DISARM’ command does not prevent subsequent measurements from being
triggered when the Test Set is in repetitive triggering mode.
If
Disarm is selected with no measurement in progress, it will have no effect on
the display.
Screens Where Field is Present
CDMA CELLULAR MOBILE RECEIVER TEST
CDMA CELLULAR MOBILE TRANSMITTER TEST
CDMA GATED POWER
Display
The top right-hand portion of the CALL CONTROL screen is used to display:
Decoded data messages received from the mobile station on the reverse control channel
or the reverse voice channel. If a decoding error occurs the raw data message bits re-
ceived from the mobile station are displayed in hexadecimal format.
Modulation quality measurements made on the mobile station’s RF carrier while on a
voice channel.
The Display field is used to select the type of mobile station information to be
displayed.
Data/Meas is displayed for AMPS measurements. NData/NMeas is
displayed for NAMPS measurements.
The :MODE command is used to Control the Display field for AMPS, TACS, and
JTACS system types.
The :NDMMode command is used to control the Display field for NAMPS system
type.
The :MODE? query is used to determine the current Display field setting for AMPS,
TACS, and JTACS system types.
The :NDMMode? query is used to determine the current Display field setting for
NAMPS system type.