P534 Pulse Width (Factory default is “100ms”)
A Totalizer mode relay is energized for a programmed duration (P534) each time the
Totalizer count (D828 / D829) is incremented. P534 controls the pulse width
(i.e. the pulse 'on' time) and is between 10 ms and 2.5 s, changeable in steps of 10 ms
Parameter P534 also defines the pulse width for a Sampler mode relay
(see “Set-up the relays” on page 69)
Note
See “Set-up the relays” on page 69 for information on Totalizer mode relays
Menu: MONITOR / [CONTROLLER /] READINGS / TOTALIZER
D828 Totalizer 1
This read-only parameter displays the Totalizer 1 count, which is the cumulative
totalized flow. To add this to the Full PV Display, see “Display configuration options” on
page 102
Set-up Totalizer 2 (daily totalized flow)
This is an 8-digit totalizer showing daily total flow (midnight to midnight) through a flow
structure. It is automatically re-set to zero at each midnight.
Note
A relay on the control unit can be configured to output a pulse for each increment to
the Totalizer Count parameter. See “Set-up the relays” on page 69 for how to do this
Menu: SETUP / [CONTROLLER /] OUTPUT / TOTALIZER
P532 Total 2 Factor (Factor default is 0.0)
One count is added to the Totalizer 2 Count (D829) for a quantity of liquid as defined by
this parameter. The display unit of measurement for D829 is defined by P533
P533 Total 2 Units (Factory default is "None")
This parameter defines the units for the Totalizer 2 Count (D829).
P534 Pulse Width (Factory default is “100 ms”)
A Totalizer mode relay is energized for a programmed duration (P534) each time the
Totalizer Count (D828 or D829) is incremented. P534 controls the pulse width
(i.e. the pulse 'on' time) and is between 10 ms and 2.5 s, changeable in steps of 10 ms
Parameter P534 also defines the pulse width for a Sampler mode relay
(see “Set-up the relays” on page 69)
Note
See “Set-up the relays” on page 69 for information on Totalizer mode relays
P536 Total 2 Source (Factory default is “MCU PV”)
Use this to select whether the control unit PV (D800), SV (D801), TV (D802) or FV
(D803) value is to be totalized over a 24 hours period for the Totalizer 2 Count (D829).
Alternatively, select None if not requiring Totalizer 2.