62
Figure 34. Improper Load Connections
DC Distribution Terminals
A single pair of terminals are designated as the positive and negative "DC Distribution Terminals" (DT's).
These two terminals may be the power supply output, the B+ at the load, or a separate pair of terminals
established expressly for distribution. Proper location of the DT's results in improved over-all performance and
reduced mutual coupling effects between separate loads using the same power supply. If remote sensing is not
used, locate the DT's as close as possible to the power supply output terminals - optimum performance will
result when the power supply output terminals themselves are used as the DT's. (See Figure 35.)
Figure 35. Location of DC Distribution Terminals Remote Sensing