Agilent Technologies Serials 2934A-01821 to 01825 3026A-01921 and above Agilent 6032A Video Gaming Accessories User Manual


 
Standard Commands for Programmable Instruments (SCPI)
100
Query Syntax STATus:OPERtion:EVENt?
Returned Parameters <NR1> (Register Value)
Examples STAT: OPER: EVEN? STATUS: OPERATIONAL: EVENT?
NTR/PTR Commands
These commands allow you to set or read the value of the Operation NTR (Negative-
Transition) and PTR (Positive-Transistion) registers. These registers serve as polarity filters
between the Operation Enable and Operation Event registers to cause the following actions:
When a bit of the Operation NTR register is set to 1 then a 1-to-0 transition of a corresponding enabled bit of Operation
Condition register will cause that bit to be set in the Operation Event register.
When a bit of the Operation PTR register is set to 1 then a 0-to-1 transition of a corresponding enabled bit of Operation
Condition register will cause that bit to be set in the Operation Event register.
If the same bits in both NTR and PTR registers are set to 1, then
any transition of that bit (if enabled) at the Operation
Condition register will set the corresponding bit in the Operation Event register.
If the same bits in both NTR and PTR registers are set to 0, then
no transition of that bit (even if enabled) at the
Operation Condition register can set the corresponding bit in the Operation Event register.
NOTE If a Operation bit is in the proper state, then programming the corresponding1 PTR or NTR filter bit to 1
will set the associated Event Register bit.
Command Syntax STATus:OPERtion:NTRansition <NRf>
STAT
us:OPERtion:PTRansition <NRf>
Parameters 0 to 32727
Default Value 0
Examples STAT:OPER:NTR32 STATUS:OPERATION:PTR 1312
Query Syntax STAT:OPER:NTR? STAT:OPER:PTR?
Returned Parameters <NR1>
(Register value)
STAT:PRES
This command sets all defined bits in the Status Subsystem PTR registers and clears all bits in the subsytem NTR and
Enable registers. STAT:OPER:PTR is set to 1313 and STAT:QUES:PTR is set to 1555.
Command Syntax STATus:PRESet
Examples STAT: PRES STATUS: PRESET
Status Questionable Registers
The bit configuration of all Status Questionable registers is as follows:
Bit Configuration of Questionable Registers
Bit Position 15-1110 9876 5 43210
Condition NU UNR RI NU NU NU NU TO NU NU OC OV
Bit Weight 1024 512 256 128 64 32 16 8 4 2 1
NU = (Not used); OC = Overcurrent protection circuit has tripped; OT = Overtemperature status condition exists;
OV = Overvoltage protection circuit has tripped; RI = Remote inhibit is active; UNR = Power supply output is
unregulated.
See "Status Reporting" for more explanation of these registers.