Agilent Technologies Serials 2934A-01821 to 01825 3026A-01921 and above Agilent 6032A Video Gaming Accessories User Manual


 
Standard Commands for Programmable Instruments (SCPI)
110
Hardware Errors During Operation. If an error does not occur during selftest but after the power supply has been operating
correctly for a time, error -240 is placed in the error queue. Most subsequent commands will not be executed and will also
cause one or more - 240 errors to be placed in the queue after each command. If your application requires the controller to
be signalled if the power supply fails, program the status registers to allow execution errors or device-dependent errors to
generate an SRQ to the controller.
Table C-5. Summary of Error Messages
Error Number Error String [Description/Explanation/Examples]
-100 Command error [generic]
-101 Invalid character
-102 Syntax error [unrecognized command or data type]
-103 Invalid separator
-104 Data type error [e.g., "numeric or string expected, got block data"]
-105 GET not allowed
-108 Parameter not allowed [too many parameters]
-109 Missing parameter [too few parameters]
-110 Command header error
-111 Header separator error
-112 Program mnemonic too long [maximum 12 characters]
-113 undefined header [operation not allowed]
-118 Query not allowed
-120 Numeric data error
-121 Invalid character in number [includes ’’9" in octal data "#q", etc.]
-123 Exponent too large [numeric overflow; exponent magnitude > 32 k]
-124 Too many digits [number too long; more than 255 digits received]
-128 Numeric data not allowed
-130 Suffix error
-131 Invalid suffix [unrecognized units, or units not appropriate]
-134 Suffix too long
-138 Suffix not allowed
-140 Character data error
-141 Invalid character data [bad character, or unrecognized]
-144 Character data too long [maximum length is 12 characters]
-148 Character data not allowed
-150 String data error
-151 Invalid string data [e.g., END received before close quote]
-158 String data not allowed
-160 Block data error
-161 Invalid block data [e.g., END received before length satisfied]
-168 Block data not allowed
-170 Expression error
-171 Invalid expression [e.g., illegal character in expression]