Agilent Technologies Serials 2934A-01821 to 01825 3026A-01921 and above Agilent 6032A Video Gaming Accessories User Manual


 
General Information
18
Table 1-2. Supplemental Characteristics (continued)
DC Floating Voltage
: Either output terminal may be floated up to the following voltage (including the output voltage) from
earth ground:
±
240 Vdc on Models 6031A, 6032A, 6033A, and 6038A
±
550 Vdc on Models 6030A and 6035A
Exceeding these voltage can result in damage to the equipment.
Remote Sensing: The power supply maintains specifications at the load with up to 0.5 volt drop per load lead. Operation
with up to 2 volts per load lead is possible with some degradation of the load effect specification.
Absolute maximum ratings
between FLT Terminals 1 and 2 is 17.5 Volts dc. (external resistor required to limit current to 1.25 mA max)
between INH Terminals 3 and 4 is 5.5 Volts dc.
between Terminals 1 or 2 to 4 or chassis to ground is 42 Volts dc.
Fault (FLT) levels (Terminals 1 and 2) (All models)
Ioh (Low Level Output Current) is + 1.2 mA maximum
Vol (Low Level Output Voltage) is 0.5 Volts Maximum
Inhibit (INH) Levels (Terminals 3 and 4) (All models)
Vih (High Level Input Voltage) is 2 Volts
Vil (Low Level Input Voltage) 0.5 Volts Maximum
tw (Pulse Width) (minimum) = 50 ms (2 ms typ)
td (Delay Time) = 2 ms typ
Iil (Low Level Input Current) = -1.25 mA maximum
Multiple Operations: Up to two similar units may be connected in series or auto-parallel, to provide increased output
capabilities. Mixing supplies with dissimilar output capabilities is not recommended because under certain conditions, the
lower output supply may be stressed beyond its maximum voltage and or current capabilities by the higher output supply.
Temperature Rating (
°
C):
Operating is 0-50 (Agilent 6030/6031/6032/6035); 0-55 (Agilent 6033/6038)
Storage is - 40 + 75 (all models)
Weight kg. (Ibs)
Model Agilent 6030A Agilent 6031A Agilent6032A Agilent 6033A Agilent 6035A Agilent 6038A
Net 16.3 (36) 17.2 (38) 16.3 (36) 9.6 (21) 16.3 (36) 9.6 (21)
Shipping 21.7 (48) 22.7 (50) 21.8 (48) 11.4 (25) 21.7 (48) 11.4 (25)
Dimensions: See Figure 2-1.
Certification:
The unit is designed to comply with these requirements:
ICE 348-Safety Requirements for Electronic Measuring Apparatus.
CSA Electrical Bulletin 556B-Electronic Instruments and Scientific Apparatus for Special Use and Applications.
VDE 0871.6.78 Level B-RFI Suppression of Radio Frequency Equipment for Industrial, Scientific, and Medical (ISM)
and similar purposes.
VDE 0411-Electronic Measuring Instruments and Automatic Controls.
UL 1244-Electrical and Electronic Measuring & Testing Equipment.
ANSI C39.5 Part 0 Draft 8-Electrical Testing, Measurement, and Control Equipment.