Initiate Subsystem
This subsystem enables the trigger system. When a trigger is enabled, the triggering action will occur upon receipt of a
<GET>,
*TRG, or TRIGger command. If a trigger circuit is not enabled, all trigger commands are ignored. The actions of
the two initiate commands are:
If
INIT:CONT is OFF, then INIT[:IMM] arms the trigger system for a single trigger.
If
INIT:CONT is ON, then the trigger system is continuously armed and INIT[:IMM] is redundant.
Command Syntax INITiate[:IMMediate]
INITiate:CONTinuous <bool>
Parameters
For INIT[:IMM] (None)
For INIT:CONT 0 or OFF 1 or ON
*RST Value OFF
Examples INIT INITIATE: IMMEDIATE
INIT:CONT 1 INITIATE:CONTINUOUS 1
Query Syntax
For INIT[:IMM] (None)
For
INIT:CONT INIT:CONT?
Returned Parameters <NR1> 0 or 1
Measure Subsystem
This query subsystem returns the voltage and current measured at the power supply’s sense
Query Syntax MEASure:CURRent[:DC?] MEASure:VOLTage[:DC?] Default Suffix
A for MEAS:CURR V for MEAS:VOLT
Examples MEAS: CURR? MEAS: VOLT?
MEASURE: VOLTAGE: DC? MV
Returned Parameters <NR3>
Output Subsystem
This subsystem controls the power supply’s voltage and current outputs and the optional Agilent 59510A/59511A Relay
Accessory.
OUTP[:STAT]
Enables or disables the power supply’s output. The exact state of a disabled output is model dependent, but it generally is in
a condition of minimum output voltage and source current. If the power supply is configured with the Agilent
59510A/59511A Relay Accessory, the command will open the relay contacts when the output is disabled and close them
when it is enabled. Use of the second (
NORelay) parameter prevents the command from having any effect on the relay; it
remains in its existing state when
OUTPut is executed. The query form returns the output state, excluding that of the relay
(see
OUTP:REL?).
Command Syntax
OUTPut[:STATe] <NRf>[,NORelay]
Parameters
{O/OFF}[,NORelay] {I/ON}[,NORelay]
*RST Value
0
Examples OUTP 1 OUTPUT: STATE ON, NORELAY
Query Syntax OUTP
ut[:STATe]?
Returned Parameters