Agilent Technologies Serials 2934A-01821 to 01825 3026A-01921 and above Agilent 6032A Video Gaming Accessories User Manual


 
Blank Front Panel Option 00180
Table B-1. ID Query Response
Agilent Model Response
Agilent 6030A ID Agilent 6030A or ID Agilent 6030A, Opt 100
Agilent 6031A ID Agilent 6031A or ID Agilent 6031A, Opt 100
Agilent 6032A ID Agilent 6032A or ID Agilent 6032A, Opt 100
Agilent 6033A ID Agilent 6033A or ID Agilent 6033A, Opt 100
If the power supply responds correctly, the microcomputer, associated circuits and software are almost certainly functioning
properly.
The most common cause of failure of the power supply to respond to commands sent from the controller is typing errors. To
determine if the power supply has detected an error, send string:
ERR?
and address the power supply to talk. The response should be:
ERR x
where x is a decimal digit from 0 to 8. See Table 3-10 for a description of error codes. If the power supply still fails to
respond, check that you are using the appropriate format for sending GP-IB commands from your controller. Check that the
GP-IB address switches on the rear panel of the power supply are set for the address being used. Substitute another GP-IB
instrument for the suspect power supply to determine that a fault exists in the power supply and not in the controller,
interface, or cables.
The following steps check the CV and CC loops and the OVP circuit. Remember that measured responses
(VOUT,IOUT,OVP) may vary within specified tolerances from the expected response.
a. To check the CV loop, send string:
VSET 20; ISET 1
and wait one second for power supply to settle. Send string:
STS?
and address the power supply to talk. The response should be:
STS 1
indicating that the power supply is in the CV mode. Send string:
VOUT?
and address the power supply to talk. The response should be;
VOUT 20.000
b. To check the OVP circuit, send string:
VSET 10
To run the following program (Agilent-85 format is used here.