Agilent Technologies Serials 2934A-01821 to 01825 3026A-01921 and above Agilent 6032A Video Gaming Accessories User Manual


 
Operating Instructions
51
Table 3-7 GP-IB Commands (continued)
Command
*Range or **Response
To Query Description
Page
DLY x
DLY xS
DLY xMS
*0--31.999 s
*0--31999 ms
Any of these commands is used to program the
delay time after a new output voltage or current is
implemented, or RST or OUT ON command is
received. During delay power supply disables CV,
CC, and OR conditions from being labeled as
faults, and disables foldback protection.
57
DLY? **DLY xx.xxx Used to read delay time setting. 57
OUT OFF
OUT 0
.Enables or disables power supply output. Power
supply remains able to implement commands
even while output is disabled
58
OUT ON
OUT 1
Initial Condition: OUT ON
OUT? **OUT 0 or
**OUT 1
Used to read OUTPUT ON/OFF setting 58
FOLD OFF,
FOLD 0
FOLD CV
FOLD 1
FOLD CC
FOLD 2
Enables or disables foldback protection which
will disable power supply output if power supply
switches to whichever mode, CV or CC, is
defined as the fold (error) condition. Foldback
protection is inhibited during DELAY period.
Initial Condition: FOLD OFF.
58
FOLD? **FOLD 0 or
**FOLD 1 or
**FOLD 2
Used to read FOLDBACK setting.
58
RST Used to reset power supply if output is disabled
by overvoltage, remote inhibit, or foldback
protection circuits. Power supply resets to present
voltage and current settings (values can be
changed while unit is disabled).
59
HOLD OFF
HOLD 0
HOLD ON HOLD
1
Determine if certain newly received commands
are implemented by power supply upon receipt or
are held for later
implementation while power supply
continues to operate with previously received
values. HOLD ON can be used to synchronize
power supply changes with actions taken by other
devices on the GP-IB. See TRG command. Initial
Condition: HOLD OFF
59
HOLD? **HOLD 0 or
**HOLD 1
Used to read HOLD setting 59