8 – Language Dictionary
96
Calibration Commands
Calibration commands let you enable and disable the calibration mode, change the calibration password,
calibrate current and voltage programming, and store new calibration constants in nonvolatile memory.
NOTE: If calibration mode has not been enabled with CALibrate:STATe, programming the
calibration commands will generate an error. You must also save any changes that you made
using CALibrate:SAVE, otherwise all changes will be lost when you exit calibration mode.
CALibrate:CURRent
This command initiates the calibration of the positive dc output current as well as the high-range current
measurement circuit.
Command Syntax
CALibrate:CURRent[:SOURce][:DC][:POSitive]
Parameters
None
Examples
CAL:STAT 1, ! enable calibration
CAL:CURR, ! start current calibration
Related Commands
CAL:CURR:NEG, CAL:LEV, CAL:DATA
CALibrate:CURRent2
Agilent 66319B/D only
This command initiates the current calibration of output 2.
Command Syntax
CALibrate:CURRent2
Parameters
None
Examples
CAL:CURR2
Related Commands
CAL:CURR:NEG, CAL:LEV, CAL:DATA
CALibrate:CURRent:MEASure:R3
This command initiates the calibration of the middle-range current measurement circuit.
Command Syntax
CALibrate:CURRent:MEASure:R3
Parameters
None
Examples
CAL:CURR:MEAS:R3
Related Commands
CAL:CURR
CALibrate:CURRent:MEASure:LOWRange
This command initiates the calibration of the low-range current measurement circuit.
Command Syntax
CALibrate:CURRent:MEASure[:DC]:LOWRange
Parameters
None
Examples
CAL:CURR:MEAS
Related Commands
CAL:CURR