Agilent Technologies 66319b Video Games User Manual


 
Language Dictionary - 8
131
Command Syntax
TRIGger:SEQuence2:SLOPe:DVM<slope>
TRIGger:ACQuire:SLOPe:DVM<slope>
Parameters
EITHer | POSitive | NEGative
*RST Value
POSitive
Examples
TRIG:SEQ2:SLOP:DVM POS TRIG:ACQ:SLOP:DVM EITH
Query Syntax
TRIGger:SEQuence2:SLOPe:DVM?
TRIGger:ACQuire:SLOPe:DVM?
Returned Parameters
<CRD>
Related Commands
TRIG:SEQ2:LEV:DVM
TRIGger:SEQuence2:SLOPe:VOLTage
TRIGger:ACQuire:SLOPe:VOLTage
This command sets the slope of an internally triggered voltage measurement.
POSitive
triggering occurs on the rising edge.
NEGative
triggering occurs on the falling edge.
EITHer
triggering occurs on either edge.
Command Syntax
TRIGger:SEQuence2:SLOPe:VOLTage<slope>
TRIGger:ACQuire:SLOPe:VOLTage<slope>
Parameters
EITHer | POSitive | NEGative
*RST Value
POSitive
Examples
TRIG:SEQ2:SLOP:VOLT POS TRIG:ACQ:SLOP:VOLT EITH
Query Syntax
TRIGger:SEQuence2:SLOPe:VOLTage?
TRIGger:ACQuire:SLOPe:VOLTage?
Returned Parameters
<CRD>
Related Commands
TRIG:SEQ2:SLOP:CURR
TRIGger:SEQuence2:SOURce
TRIGger:ACQuire:SOURce
These commands select the trigger source for measurement triggers as follows:
BUS
GPIB device, *TRG, or <GET> (Group Execute Trigger)
INTernal
trigger is generated internally when the measured waveform crosses the trigger
level with the selected slope.
EXTernal
Selects the external trigger input on the 4-pin control connector as the trigger
source (only applies to units with firmware revision A.03.01 and up)
Command Syntax
TRIGger:SEQuence2:SOURce<source>
TRIGger:ACQuire:SOURce<source>
Parameters
BUS | INTernal
*RST Value
INTernal
Examples
TRIG:ACQ:SOUR BUS
Query Syntax
TRIGger:SEQuence2:SOURce?
TRIGger:ACQuire:SOURce?
Returned Parameters
<CRD>