Agilent Technologies 66319b Video Games User Manual


 
5 – Front Panel Operation
60
10 - Storing and Recalling Instrument States
You can save up to 4 states (from location 0 to location 3) in non-volatile memory and recall them from
the front panel. All programmable settings are saved. This capability is only available when the unit is set
to the SCPI programming language.
NOTE: You can program the unit to automatically power-on according to the instrument state
that is saved in state 0 as shown in the third example.
Save an instrument state in location 0 as follows:
Action Display
1. Set the instrument to the state that you want to save.
2.
Save this state to location 0. Press Save, Enter Number, 0, Enter.
*SAV 0
Recall a saved state as follows:
Action Display
1.
Recall the state saved in location 0 by pressing Recall, Enter Number, 0, Enter
*RCL 0
Select the power-on state of the dc source as follows:
Action Display
1.
On the Function keypad, press Output, and scroll through the Output menu until you
get to the PON state command.
PON:STATE RST
2. Use the # and & keys to select either RST or RCL0. RST sets the power-on state of
the unit as defined by the *RST command. RCL0 sets the power-on state of the unit to
the state saved in *RCL location 0. Press Enter when done.
Clear the non-volatile memory of the dc source as follows:
Action Display
1.
On the Function keypad, press Output, Enter. This returns the unit to the factory-
default settings.
*RST
2.
Save these settings to location 0. Press Save, Enter Number, 0, Enter.
*SAV 0
3. Repeat step #2 for memory locations 1 through 3. *SAV 1
*SAV 2
*SAV 3