TRIGger:SEQuence2:LEVel:VOLTage
TRIGger:ACQuire:LEVel:VOLTage
This command sets the trigger level for internally triggered voltage measurements. A positive voltage
trigger occurs when the voltage level changes from a value less than the lower hysteresis band limit to a
value greater than the upper hysteresis band limit. Similarly, a negative voltage trigger occurs when the
voltage level changes from a value greater than the upper hysteresis band limit to a value less than the
lower hysteresis band limit.
Command Syntax
TRIGger:SEQuence2:LEVel:VOLTage<NRf+>
TRIGger:ACQuire:LEVel:VOLTage<NRf+>
Parameters
0 to MAX (see table 8-3)
Unit
V (volts)
*RST Value
0
Examples
TRIG:SEQ2:LEV:VOLT 5
TRIG:ACQ:LEV:VOLT MAX
Query Syntax
TRIGger:SEQuence2:LEVel:VOLTage?
TRIGger:ACQuire:LEVel:VOLTage?
Returned Parameters
<NR3>
Related Commands
TRIG:SEQ2:LEV:CURR TRIG:SEQ2:HYST:VOLT
TRIGger:SEQuence2:SLOPe:CURRent
TRIGger:ACQuire:SLOPe:CURRent
This command sets the slope of an internally triggered current measurement.
POSitive
triggering occurs on the rising edge.
NEGative
triggering occurs on the falling edge.
EITHer
triggering occurs on either edge.
Command Syntax
TRIGger:SEQuence2:SLOPe:CURRent<slope>
TRIGger:ACQuire:SLOPe:CURRent<slope>
Parameters
EITHer | POSitive | NEGative
*RST Value
POSitive
Examples
TRIG:SEQ2:SLOP:CURR POS TRIG:ACQ:SLOP:CURR EITH
Query Syntax
TRIGger:SEQuence2:SLOPe:CURRent?
TRIGger:ACQuire:SLOPe:CURRent?
Returned Parameters
<CRD>
Related Commands
TRIG:SEQ2:SLOP:VOLT
TRIGger:SEQuence2:SLOPe:DVM
TRIGger:ACQuire:SLOPe:DVM
Agilent 66321D/66319D only
This command sets the slope of the DVM input signal.
POSitive
measurement triggering occurs on the rising edge.
NEGative
measurement triggering occurs on the falling edge.
EITHer
measurement triggering occurs on either edge.