Agilent Technologies 66319b Video Games User Manual


 
Language Dictionary - 8
105
MEASure:VOLTage2
Agilent 66319B/D only
This query measures the output voltage at the auxiliary output. Output 2 measurements are calculated
from a total of 2048 readings taken at a 15.6 microsecond sampling rate. These parameters are fixed.
Query Syntax
MEASure[:SCALar]:VOLTage2[:DC]?
Parameters
None
Examples
MEAS:VOLT2? FETC:VOLT2:DC?
Returned Parameters
<NR3>
Related Commands
MEAS:CURR2?
MEASure:VOLTage:ACDC?
FETCh:VOLTage:ACDC?
These queries return the ac+dc rms output voltage.
Query Syntax
MEASure[:SCALar]:VOLTage:ACDC?
FETCh[:SCALar]:VOLTage:ACDC?
Parameters
None
Examples
MEAS:VOLT:ACDC? FETC:VOLT:ACDC?
Returned Parameters
<NR3>
Related Commands
MEAS:CURR:ACDC?
MEASure:VOLTage:HIGH?
FETCh:VOLTage:HIGH?
These queries return the High level voltage of a voltage pulse waveform. The instrument first measures
the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse
waveform using 16 bins between the maximum and minimum data points. The bin containing the most
data points above the 50% point is the high bin. The average of all the data points in the high bin is
returned as the High level. If no high bin contains more than 1.25% of the total number of acquired
points, then the maximum value is returned by these queries.
Query Syntax
MEASure[:SCALar]:VOLTage:HIGH?
FETCh[:SCALar]:VOLTage:HIGH?
Parameters
None
Examples
MEAS:VOLT:HIGH? FETC:VOLT:HIGH?
Returned Parameters
<NR3>
Related Commands
MEAS:VOLT:LOW? CALC:REF:HIGH