Agilent Technologies 66319b Video Games User Manual


 
Language Dictionary - 8
109
SENSe:SWEep:POINts
This command defines the number of points in a measurement.
Command Syntax
SENSe:SWEep:POINts<NRf+>
Parameters
1 through 4096
*RST Value
2048
Examples
SENS:SWE:POIN 1024
Query Syntax
SENSe:SWEep:POINts?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR
SENSe:SWEep:TINTerval
This command defines the time period between samples. The value that you enter for the time interval
will be rounded to the nearest 15.6 microsecond increment.
Command Syntax
SENSe:SWEep:TINTerval<NRf+>
Parameters
15.6 microseconds through 31200 seconds
*RST Value
15.6 microseconds
Examples
SENS:SWE:TINT 31.2E-6
Query Syntax
SENSe:SWEep:TINTerval?
Returned Parameters
<NR3>
Related Commands
SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR
SENSe:WINDow
This command sets the window function that is used in dc and in ac+dc rms measurement calculations.
The following functions can be selected:
HANNing
A signal conditioning window that reduces errors in dc and rms measurement
calculations in the presence of periodic signals such as line ripple. It also reduces
jitter when measuring successive pulse waveforms. The Hanning window multiplies
each point in the measurement sample by the function cosine
4
. Do not use the
Hanning window when measuring single-shot pulse waveforms.
RECTangular
A window that returns measurement calculations without any signal conditioning.
This window may be used for pulse measurements where the exact period of the
pulse waveform is known and the measurement interval can be set accordingly using
the SENSe:SWEep:TINTerval command.
NOTE: Neither window function alters the instantaneous voltage or current data returned in the
measurement array.
Command Syntax
SENSe:WINDow[:TYPE] <type>
Parameters
HANNing | RECTangular
*RST Value
HANNing
Examples
SENS:WIND RECT
Query Syntax
SENSe:WINDow[:TYPE]?
Returned Parameters
<CRD>