Agilent Technologies 66319b Video Games User Manual


 
Performance, Calibration, and Configuration - B
157
Front Panel Calibration Procedure
These procedures assume you understand how to operate front panel keys (see chapter 5). Make sure the
sense terminals are directly jumpered to the output terminals.
Enable Calibration Mode
Action Display
1.
Reset the unit by selecting Output, scrolling to *RST and pressing Enter.
*RST
2.
Press Output On/Off to enable the output.
00.003V 0.0006A
3.
To begin calibration press Shift Cal, scroll to CAL ON and press Enter.
CAL ON 0.0
4.
Enter the calibration password from Entry keypad and press Enter. lf the
password is correct the Cal annunciator will come on.
If CAL DENIED appears, then an internal switch has been set to prevent the
calibration from being changed.
lf the password is incorrect, an error occurs. If the active password is lost, the
calibration function can be recovered by setting an internal switch that defeats
password protection (contact the factory for details).
CAL DENIED
OUT OF RANGE
Voltage Programming and Measurement Calibration
Action Display
5. Connect the external DMM (in voltage mode) directly to output 1.
Do not connect the load resistor or current shunt. Select output 1.
6.
Press Shift Cal, scroll to CAL VOLT, and press Enter.
CAL:VOLT
7.
Press Shift Cal, scroll to CAL LEV, and press Enter to select the first
calibration point.
CAL:LEV P1
8.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter the
voltage value displayed on the DMM. Press Enter.
CAL:DATA 0.00
9.
Press Shift Cal, scroll to CAL LEV, use & to scroll to P2 (the second
calibration point), and press Enter.
CAL:LEV P2
10.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter the
second voltage value displayed on the DMM. Press Enter.
CAL:DATA 0.00
Agilent 66319B/D Output 2 Voltage Programming and Measurement Calibration
Action Display
11. Connect the external DMM (in voltage mode) directly to output 2.
Do not connect the load resistor or current shunt. Select output 2.
12.
Press Shift Cal, scroll to CAL VOLT2, and press Enter.
CAL:VOLT2
13.
Press Shift Cal, scroll to CAL LEV, and press Enter to select the first
calibration point.
CAL:LEV P1
14.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter the
voltage value displayed on the DMM. Press Enter.
CAL:DATA 0.00
15.
Press Shift Cal, scroll to CAL LEV, use & to scroll to P2 (the second
calibration point), and press Enter.
CAL:LEV P2
16.
Press Shift Cal, scroll to CAL DATA, press Enter Number, and enter the
second voltage value displayed on the DMM. Press Enter.
CAL:DATA 0.00