Agilent Technologies 71501D Video Eyeware User Manual


 
1-25
Getting Started
Steps for Setting Up Eye-Diagram Analysis
70841A/B appears.
c Press ADJUST ROW, and rotate the knob to move the box to the row where
the 70841A/B appears.
d Press ASSIGN BOTH.
4 Use the displayed softkey menus to set the pattern generator to the desired
settings. For an example, refer to the paragraph "Configure the data signal" in
Chapter 2, Application Tutorials.
5 To return control to the eye-diagram analyzer, press:
DISPLAY, NEXT INSTR
6 To return to the eye-diagram analyzer personality menus, press:
USER
To Manually Set
the Trigger Level
If a pattern generator other than the 70841A/B is used, you must use the fol-
lowing procedure to manually set the trigger level.
1 Select trace two and turn the display on by pressing:
> prev menu, select:, TR2, display ON/OFF ON
2 Automatically scale the trigger trace and set the trigger level by pressing:
AUTO-SCALE, Trg,Cal, LEVEL
Use the knob to position the trigger level indicator line so it crosses an edge.