Agilent Technologies 71501D Video Eyeware User Manual


 
Eye-Diagram Analyzer Reference
Calibrating the Eye-Diagram Analyzer
1 Connect a low-loss cable between the front-panel CALIBRATOR OUTPUT and
channel 1 connectors.
2 To calibrate the IF, press:
MENU, page 1 of 2, Calib, IF calib, CAL CH1
The calibration routine takes about 30 minutes to complete.
3 Disconnect the cable from channel 1, and connect it to the channel 2
connector.
4 Repeat the above procedure for channel 2.