Agilent Technologies 71501D Video Eyeware User Manual


 
Eye-Diagram Analyzer Reference
Agilent 70820A User-Corrections
To Delete a
Channels User-
Corrections
1 Delete a channels user-corrections by pressing:
MENU, page 1 of 2, Calib, user corr
2 Select the desired channel by pressing:
CH1|CH2
3 Delete the selected channels user-corrections by pressing:
other, DELETE ALL
4 To delete all corrections, press:
CONFIRM DEL ALL
5 To cancel the deletion process, press:
ABORT DEL ALL
Entering and Editing User-Corrections
The User Correction menu provides softkeys for entering and editing lines of
user-correction data. To view user-correction data graphically at the same
time it is created, refer to Viewing User-Correction Data on page 3-56.
The SEL|EDT softkey determines whether the front-panel knob and arrow
keys select a field for editing the current fields data. New lines can be entered
between the first and last lines. Refer to "To enter a line of data" in this sec-
tion.
The microwave transition analyzer interpolates the data between adjacent
points. Interpolation between the defined points can be either slope or flat
line. If slope is selected, the slope LOG|LIN softkey determines whether
slopes are linear or logarithmic. To locate this softkey, press other in the User
Correction menu.