Agilent Technologies 71501D Video Eyeware User Manual


 
Eye-Diagram Analyzer Reference
Agilent 70820A Calibration
Trace 2 Delay Compensated
To Delay a
Channel
Automatically compensate channel 2 by pressing:
MENU, page 1 of 2, Calib, chan skew, AUTO SKEW
Calibrating the IF
The microwave transition analyzer module provides separate automatic IF cal-
ibration routines for channel 1 and channel 2. Each requires approximately 30
minutes to run. For optimum performance, you should perform the calibra-
tions when at least one of the following is true:
The temperature has changed ±5° Centigrade since the last calibration was
performed.
More than one week has passed since the last calibration.
Before any critical measurement.