Agilent Technologies 71501D Video Eyeware User Manual


 
3-55
Eye-Diagram Analyzer Reference
Agilent 70820A User-Corrections
To Edit a Line of
Data
1 To edit a line of data, press:
MENU, page 1 of 2, Calib, user corr, SEL|EDT EDT
Use the front-panel knob, step keys, or numeric keypad to change the dis-
played value.
2 If you are editing the path to next FREQ field, change the value by pressing:
SLOPE| FLAT
To Delete a Line of
Data
1 Select the line of data to be deleted by pressing:
MENU, page 1 of 2, Calib, user corr, SEL|EDT SEL
Use the front-panel knob or arrow keys to highlight any field in the line you
wish to delete.
2 Delete the highlighted text by pressing:
other, DELETE LINE
To Select Flat-
Line Interpolation
1 Highlight the desired field by pressing:
MENU, page 1 of 2, Calib, user corr, SEL|EDT SEL
Use the front-panel knob or arrow keys to highlight the desired path to next
FREQ field.
2 Select flat-line interpolation by pressing:
SLOPE| FLAT FLAT
To Select
Logarithmic Slope
Interpolation
Select logarithmic slope interpolation by pressing:
MENU, page 1 of 2, Calib, user corr, other, slope LOG|LIN LOG
For logarithmic interpolation to take effect, the path to next FREQ field in the
lines of data must be set to slope.