Agilent Technologies 71501D Video Eyeware User Manual


 
4-5
Programming Commands
Introduction
BIT
The BIT command string sets the instruments time scale (seconds-per-divi-
sion) to a value which corresponds to the number of bits specified in the
Param variable.
Example
OUTPUT Mta;"PROG:NUMB Param,3.5"
OUTPUT Mta;"PROG:STR Command,BIT"
OUTPUT Mta;"PROG:WAIT?"
ENTER Mta;A$
OUTPUT Mta;"PROG:STR? Response"
ENTER Mta;Response$
OUTPUT Mta;"PROG:STAT CONT"
CLEAR
The CLEAR command string clears the display for persistence mode or error-
trace mode. CLEAR also restarts any trace averaging.
Values of Program Variables
Variable Data Type Value
Command
string BIT
Param
floating-point number example: 3.5
Response
string DONE
Values of Program Variables
Variable Data Type Value
Command
string CLEAR
Param
There are no parameters for this command.
Response
string DONE