Agilent Technologies 6050A Video Games User Manual


 
Local Operation 51
Table 4-1. Controls and Indicators (continued)
Item Description
6 CHAN Keys
Used in conjunction with the ENTRY keys to select a channel (module) for front panel
control and/or display.
- Identifies which module is installed in the selected input channel.
and Increment( ) and decrement ( ) the channel number.
7 FUNCTION Keys
- Returns the display to the metering function. With the metering function selected, the
display will show the measured input voltage and current, the computed input power, or certain
status conditions (e.g. INPUT SHORT ON, OC, etc.) for the selected channel. Press the Meter
key to continually step through the displays.
- Displays the selected channel’s range setting for current (C:RNG) or resistance
(R:RNG), depending upon which function is selected. The settings can be changed using the
ENTRY keys.
- Toggles the selected channel’s input on and off. Input Off disables the input. Input
On enables the input and returns the module to the original settings.
- Toggles the selected channel’s short circuit mode on and off. Short On applies a
short circuit across the input. Short Off removes the short circuit and returns the channel to the
original settings. Short On is overridden by Input Off.
- Toggles transient operation on and off for the selected channel. The Tran
annunciator is on while transient operation is on. Transient operation causes the module’s input to
periodically switch between two levels.
- Displays the selected channel’s transient level for current (C:TLV), resistance
(R:TLV), or voltage (V:TLV) depending upon which function is selected. This level can be
changed using the ENTRY keys. The input alternates between the transient level (TLV) and the
main level of the active mode (CURR, RES, or VOLT) when transient operation is turned on.
- (shifted Tran Level key) - Displays the selected channel’s slew setting for current
(C:SLW) or voltage (V:SLW) depending upon which function is selected. The settings can be
changed using the ENTRY keys. The slew settings determine the rates at which new programmed
values will change. Note that resistance changes use the voltage or current slew rate settings
depending upon the resistance range.
- Displays the frequency setting of the transient generator (e.g. FREQ 1000) for the
selected channel. The setting can be changed using the ENTRY keys. The Freq setting
determines the frequency in continuous transient operation.
(shifted Freq key) - Displays the duty cycle of the transient generator (e.g. DCYCLE
50.0) for the selected channel. The setting can be changed using the ENTRY keys. The Dcycle
setting determines the TLEV portion (percentage) of the duty cycle in continuous transient
operation.
Clears the selected channel’s latching-type protection circuits: overvoltage, overpower,
overtemperature, and overcurrent (user programmed).