Agilent Technologies 5962-8200 Video Gaming Accessories User Manual


 
3 - Troubleshooting
42
Line Voltage Wiring
Figure 3-3 illustrates the primary wiring configuration of the power transformer for various ac line voltages.
Use long nose pliers to disconnect the wires going to the transformer terminals.
NOTE: Install the correct fuse when changing the ac line voltage from a previous setting:
for 110/120 Vac: 2.5AT, 250V, Agilent p/n 2110-0633;
for 220/230 Vac: 1.25AT, 250V, Agilent p/n 2110-0788
Figure 3-3. Transformer Wiring
white/red/grey
grey
orange
orange
1
2
3
4
5
6
7
Top part of
transformer
Front of unit
white/red/grey
grey
orange
white/red/grey
grey
orange
orange
white/red/grey
grey
orange
220 VAC
(spare)
orange
(spare)
orange
white/red
white/black
white/brown
white/brown
white/black
white/red
red
black
white/yellow
white/violet
white/yellow
white/violet
white/yellow
white/violet
white/yellow
white/violet
Top part of
transformer
Front of unit
120 VAC
Top part of
transformer
Front of unit
100 VAC
Top part of
transformer
Front of unit
230 VAC
1
2
3
4
5
6
7
1
2
3
4
5
6
7
1
2
3
4
5
6
7
Bottom part of
transformer
Front of unit
All Voltages